Abstract
Purpose
We report a case of significant air embolization to the ascending aorta immediately following deployment of EndoAnchors in the aortic arch during a procedure to correct a type 1A endoleak.
Case report
The novel Heli-Fx EndoAnchor system (Medtronic Vascular, Santa Rosa, CA, USA) was used to deploy helical anchors in the distal aortic arch during a procedure to correct a type 1A endoleak following Zone 2 thoracic endovascular aortic repair of a saccular proximal descending thoracic aorta aneurysm (DTAA). The patient developed ST-segment elevations principally in the inferior leads and severe hypotension moments after EndoAnchor deployment at the proximal edge of the endograft. Transesophageal echocardiogram revealed severe right ventricular hypokinesis and a large amount of air in the ascending aorta. Subsequent management and clinical and radiological 30-day follow-up is presented in addition to a review of the literature and ex vivo testing with the Heli-Fx system to examine potential causes and solutions.
Conclusion
Precautions, such as pressurized saline infusion to the side port of guiding sheath, should be used whenever manipulating catheters and sheaths such as the EndoAnchor system in the aortic arch to prevent this potentially lethal complication.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.