Abstract
Purpose:
Maldeployment of the Supera stent system can result in primary technical failure, inferior primary patency, and poorer patient outcomes. The purpose of this article is to present a case series illustrating the conditions required to perform optimal stent deployment, and if necessary, undertake successful stent removal following maldeployment.
Technique:
Two key failures of effective Supera deployment are elongation and invagination. Several technical factors should be considered to reduce the risk of maldeployment: aggressive target vessel predilation, the use of multiple fluoroscopic views, slow deployment with controlled forward pressure applied on the delivery device, “sandwich packing” of the stent above and below target lesions, and the “pulling back” of invagination. To successfully retrieve a partially deployed stent, 3 factors should be considered: the percentage of the stent already deployed, the distance from the distal tip of the introducing sheath to the proximal extent of the deployed stent, and the severity of proximal vessel disease. The higher these factors, the greater the risk of stent detachment and failed retrieval.
Conclusion:
In this series of 6 cases of maldeployment, the removal of a partially deployed Supera stent appeared to be feasible and safe, with success dependent on selected technical and anatomical considerations.
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.