Abstract
Introduction:
Modular designs enhance accuracy of limb length and offset during total hip arthroplasty. However, such interfaces can be the site of corrosion and implant failure. The purpose of this study was to review 5 cases of catastrophic femoral head trunnion dissociation to better understand clinical and implant related risk factors for this complication.
Methods:
We describe 5 cases of femoral head dissociation following significant trunnion wear. Patient and implant related factors are reviewed. Surface wear patterns were assessed using electron microscopy and wear patterns were described qualitatively.
Results:
Commonalties among 5 presented cases include male, age over 65 and increased BMI. All femoral heads were cobalt chromium with a head size of at least 36 mm or greater. 4 out of 5 had high offset. All stems were the same design. Surface wear analysis showed material removed initially on the inferomedial aspect of the trunnion where the femoral head contacted the stem.
Conclusions:
Femoral head dissociation secondary to failure of the trunnion is a rare complication. We show associations with older age, increased BMI and male patients with increased femoral head size and offset. Material is removed initially inferomedially and progresses to fracturing the superior aspect of the trunnion.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.