We report six cases of fracture of the metal backing of the CLS expansion cup. The aim of this paper is to draw attention to this mode of failure and recommend that in cases of unexplained hip pain in those with a CLS acetabular component X-rays in at least two planes are performed.
SpotornoL., RomagnoliS., IvaldoN.. The CLS System. Theoretical concept and results.Acta Orthop Belg1993; 59: 144–8.
2.
MorscherE., BerliB., JockersW., SchenkR.. Rationale of a flexible press fit cup in total hip replacement, 5 year follow-up in 280 procedures.Clin Orthop1997; 341: 42–50.
LatimerH.A., LaciewiczP.F.. Porous coated acetabular components with screw fixation, 5–10 year results.J Bone Joint Surg (Am)1996; 78-A: 975–81.
5.
TomkinsG.S., JacobsJ.J., KullL.R., RosenbergA.G., GalanteJ.O.. Primary total hip arthroplasty with a porous coated acetabular component, 7–10 year results.J Bone Joint Surg (Am)1997; 79-A: 169–76.
6.
EnghC.A.Jr., CulpepperW.I., EnghC.A.. Long-term results of use of the anatomic medullary locking prosthesis in total hip arthroplasty.J Bone Joint Surg (Am)1997; 79-A: 177–84.
7.
MorscherE.W.. Endoprosthetics.Berlin, Heidleberg, New York: Springer Verlag, 1995; 271–96.
8.
BlahaJ.D., SpotornoL., RomagnoliS.. CLS press-fit total hip arthroplasty.Tech Orthop1991; 6: 80–6.
9.
CallaghanJ.J., KimY.K., BrownT.D., PedersenDR., JonsonR.C.. Concerns and improvements with uncemented metal-backed acetabular components.Clin Orthop1995; 311: 76–84.
10.
RomagnoliS.. Press-fit hip arthroplasty. A European alternative.J Arthroplasty2002; 17(4 suppl 1): S108–12.
RobinsonR.P., LovellT.P., GreenT.M.. Hip arthroplasty using the cementless CLS stem. A 2–4 year experience.J Arthroplasty1994; 9: 177–92.
13.
GrappioloG., BlahaJ.D., GruenT.A., BurasteroG., SpotornoL.. Primary total hip arthroplasty using a gritblasted, press-fit femoral prosthesis. Long-term results with survivorship analysis.Hip International2002; 12: 55–73.
14.
The Swedish National Hip Arthroplasty Register2002.
15.
The CLS system - A 16-year review.An emerging gold standard for uncemented hip replacement?Int Orthop2000; 24: Insert.
16.
DaviesM.S., ParkerB.C., WardD.A., HemE., HuaJ., WalkerP.S.. Migration of the uncemented CLS acetabular component.Orthopedics1999; 22: 219–22.
17.
RadcliffeG.S., WilsonJ.L., AshfordR.U., DeBoerP.G.. Hip replacement in patients younger than 65 years: Results of the CLS prosthesis.Hip International2003; 13: 133–41.
18.
RozkydalZ., JanicekP., DeduchJ., HudecekF.. Complications of the cup of CLS Total Hip Replacement.Acta Chir Orthop Traumatol Cechosl2001; 68: 85–92.
19.
LeeP.C., ShihC.H., ChenW.J., TuY.K., TaiC.L.. Early polyethylene wear and osteolysis in cementless total hip arthroplasty: the influence of femoral head size and polyethylene thickness.J Arthroplasty1999; 14: 976–81.
20.
LivermoreJ., IlstrupD., MorreyB.. Effect of femoral head size on wear of polyethylene acetabular component.J Bone Joint Surg (Am)1990; 72-A: 518–28.
21.
Rama MohanN., GrigorisP., HamblenD.L.. Failure of an uncemented press-fit acetabular cup in primary total hip arthroplasty.Hip International2002; 4: 371–77.
22.
DoreyF.J., McKellopH.A.. The long-term evaluation of total hip arthroplasty. The need for survivorship analysis. In: FinermanG.A.M.Jr., ed. Total Hip Arthroplasty Outcomes. Appendix 275–81, Churchill-Livingstone, 1998.