Abstract
Purpose
To evaluate the rotational stability and capsular bag performance of a blue light-absorbing hydrophobic acrylic open-loop single-piece intraocular lens (IOL) with a blast-finished anchor wing haptic design during the first 6 postoperative months.
Methods
In this prospective clinical study, patients with age-related cataract and potential postoperative corrected distance visual acuity (CDVA) of 0.2 logMAR or better were included. The patients received a non-toric hydrophobic acrylic single-piece IOL with an axis mark (YST0.00; Nidek Co. Ltd, Japan). Surgeries were video-taped. Retroillumination images were acquired 1 h, 1 week and 6 months postoperatively. Rotational stability was assessed by precise image overlay. At 6 months, Purkinje meter measurements were performed to evaluate tilt and decentration.
Results
In total, 100 eyes of 77 patients were included in the analysis. Mean absolute rotation was 2.1 ± 1.7° (median 1.7, range: 0–7.9) at 6 months (1 h – 6 months postoperatively). IOL rotation was ≤3° and ≤6° in 74 (74%) and 98 (98%) eyes, respectively. Mean absolute IOL rotation from the end of surgery to 6 months was 2.5 ± 2.2° (median 2.3, range: 0–15.6; n = 78). Mean tilt (pupillary axis) and decentration were 4.1 ± 1.9° (median 4.0, range: 0.5–8.2) and 0.35 ± 0.17 mm (median 0.32, range: 0.06–0.91) respectively (n = 84). Postoperatively, 98 (98%) eyes achieved a CDVA of 0.2 logMAR or better, 95 (95%) of ≤0.1 and 81 (81%) of ≤0.0.
Conclusions
This hydrophobic acrylic single-piece IOL showed an excellent rotational stability and capsular bag position with low tilt and decentration values.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.