Abstract
Purpose:
To report a case of fully dislocated XEN Gel Implant device into the anterior chamber, 18 months after its implantation and its impact on endothelial cell density.
Result:
A 75-year-old man with a history of multiple glaucoma surgeries during the last three years presented with an entirely dislocated XEN Gel implant into the anterior chamber. An endothelial cell density reduction was observed over 18 months. Thus, the implant was removed. In the short-term after surgery, the endothelial cell loss rate reduced.
Conclusion:
XEN Gel Implant can dislocate into the anterior chamber, increasing endothelial cell loss in an eye already underwent multiple surgical procedures with subsequent adverse events. Therefore, if the implant seems displaced, the endothelial cell density should be monitored, and the length of the free-tube segment within the anterior chamber should be measured, during the patient’s follow-up.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.