Abstract
Background:
Important aspects on the diagnostics of isolated gastrocnemius contractures (IGCs) have been poorly described. This study was designed to validate a new ankle range of motion (ROM) measuring device for diagnosing an IGC. In addition, we wanted to investigate the reliability of the clinical Silfverskiöld test.
Methods:
Twelve health care personnel (24 feet) were examined by 4 testers on 3 different occasions for the reliability testing of the new ankle ROM measuring device. The same participants were examined using the Silfverskiöld test to examine the reliability of the clinical test. Eleven patients (15 feet) with IGC were examined before gastrocnemius recession, immediately after surgery, and 3 months after surgery to examine the validity and responsiveness of the ankle ROM device.
Results:
An intraclass correlation coefficient (ICC) >0.85 was found for both inter- and intrarater reliability for the new ankle ROM device. The device confirmed an IGC in 13 of 15 feet before surgery and 3 of 13 feet at 3-month follow-up. At baseline, the measured ankle dorsiflexion was median 3 degrees with the knee in extension, which increased to 10 degrees (
Conclusion:
The new ankle ROM measuring device was reliable and responsive for detecting IGC. The Silfverskiöld test had poor inter- and intrarater reliability.
Level of evidence
Level II, prospective cohort study.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.