Abstract
Aim
To evaluate the feasibility of robotically enhanced preparation of internal mammary arteries (IMA).
Method
Via three trocars in left thoracic wall the left, right or both IMA were skeletonized under CO2 insufflation and single lung ventilation using electrocautery.
Results
In 12 months, 26 LIMA, five BIMA and one RIMA were dissected. In five patients, the procedure had to be determined (IMA injury (two), respiratory insufficiency (two), and heart penetration (one)). Mean intrathoracic pressure was 9.7 ± 1.5 mmHg. Mean time for LIMA and RIMA dissection was 66.7 ± 21.1 and 99.2 ± 8.7 min, respectively. In 10 patients, pericardium was incised and course of LAD assessed. However, in two patients, this coronary did not correlate with LAD. Time for instrument change depended on type of tool (cautery blade: 24.9 ± 13.1 s, clip applier 72.8 ± 28.4 s).
Conclusion
Robotic dissection of IMA is reasonable. However, life-threatening complications can barely be managed due to inadequate tools and excessive time for instrument change. Incorrect determination of coronaries can result in misplaced anastomoses.
Get full access to this article
View all access options for this article.