Abstract
Medication infusion pumps are the most popular device in almost all areas of a hospital; therefore, it is important to frequently inspect the accuracy of the infusion pump operation to prevent underdose/overdose accidents. However, the conventional infusion pump inspection devices are not suitable for quick and convenient on-site inspection by nurses. In this study, a new IR estimation technique for peristaltic infusion pumps that facilitates on-site pre-screening test with shorter inspection time was proposed. A thin membrane potentiometer was attached to a catheter and the actual IR was estimated based on a time interval between two successive line pushes of an identical cam follower using power function estimation. To evaluate the performance of the proposed IR estimation technique, in vitro experiments were performed using 11 infusion pumps (three for Infusion Pump SET 1 (