Abstract
The longevity of total hip arthroplasty (THA) continues to improve with advancements in design and bearing materials. However, the incidence of dislocation and impingement-related failures continue to rise, with the inability of the surgeon to achieve optimal component orientation cited as a cause. Computer-assistance has been shown to increase the accuracy of component orientation and robotic-assistance has been developed to translate this advantage into precise surgical execution. We sought to validate a haptically-guided robotic arm system in performing THA with the aim of comparing the accuracy of robotic-assisted acetabular cup placement to manual placement. We implanted 12 acetabular components in 6 cadaveric pelvises comparing robotic-assistance on one side with manual implantation on the other. We measured planned and actual center of rotation (COR), cup position, leg-length equalization and offset for each THA using computed tomography and the robotic platform. The root-mean-square (RMS) error for the robotic-assisted system was within 3° for cup placement and within 1mm for leg-length equalization and offset when compared to computed tomography. The robotic-assisted system was significantly more accurate than manual implantation in reproducing the COR and cup orientation, as determined by a preoperative plan. The RMS error for manual implantation compared to robotic-assistance was 5 times higher for cup inclination and 3.4 times higher for cup anteversion (