Abstract
In vitro wear data for hip joint devices reported in the literature vary in a wide range from one simulator study to another sometimes for the same type of device tested under identical physiological testing conditions. We hypothesized that non-bearing surface condition of the testing components could be an important factor affecting the simulator wear results. To confirm this hypothesis, fifteen 50 mm metal-on-metal hip resurfacing devices with identical bearing specifications were tested in a ProSim hip wear simulator for 5 million cycles. The heads were standard Birmingham Hip Resurfacing (BHR) heads; whilst the pairing acetabular cups were identical to the standard BHR cup except their different back surface conditions, including: (a) off-the-shelf products after removing the hydroxyapatite (HA) coating; (b) semi-finished products without HA coating; and (c) purposely-made cups without cast-in beads and HA coating. Results showed that the different back surfaces of the cups used indeed caused significantly large variations in the gravimetrically measured wear loss. We postulated that materials loss from the non-bearing surface of the testing components could contribute to the gravimetrically measured wear loss during a wear simulator test both directly and indirectly. The results presented in this paper pertain to In vitro wear simulator study and have little clinical relevance to the performance of any implant in vivo.