Abstract
Background
The most widely used peritoneal function test, the peritoneal equilibration test (PET), is performed with a 2.27% glucose solution. Recently, the International Society for Peritoneal Dialysis committee on ultrafiltration failure (UFF) advised performing the test with 3.86% glucose solution because it is more sensitive for detecting clinically significant UFF. Because no reference values for this test were available, we analyzed the results of standard peritoneal permeability analyses (SPAs) using 3.86% glucose.
Methods
The tests were performed in our center on 154 clinically stable peritoneal dialysis (PD) patients that were free of peritonitis for at least 4 weeks. For the assessment of reference values, we used two approaches. In approach A, patients with UFF, defined as net ultrafiltration (UF)< 400 mL/4 hours, were excluded. In approach B, only patients within their first 2 years of PD treatment were included, regardless of net UF. Means and 95% confidence intervals (95% CI) were calculated for the transport parameters of the PET and SPA.
Results
Means of normal distribution with 95% CI in approach A were as follows: for 2.0-L exchanges, mass transfer area coefficient (MTAC) for creatinine 8.8 mL/minute (4.7 – 12.7 mL/min), dialysate/plasma ratio (D/P) creatinine 0.70 (0.52 – 0.88), glucose absorption 58% (44% – 72%), dialysate240/initial dialysate ratio of glucose (D
Conclusions
A peritoneal transport function test using 3.86% glucose provides data on various aspects of transport. This study gives normal reference values that can be used for analysis of causes of UFF.