The modifications of mechanical properties produced in valvular tissue by glutaraldehyde fixation were studied. Porcine aortic leaflets, fresh and fixed with glutaraldehyde in a stressed state, were tested with a mechanical device. The results we obtained show that the fixed leaflets have a failure load higher than fresh ones and an increased elastic recovery. The time of preservation (up to one year) does not affect the mechanical properties of fixed samples. The temperature of shrinkage was also studied, and we found that fixed leaflets show no shrinkage at all.
BroomN.D.: Simultaneous morphological and stress-strain studies of the fibrous components in wet heart valve leaflet tissue.Connect. Tissue Res.6, 37, 1978.
4.
CarpentierA., LemaigreG., RobertL., CarpentierS., and DubostC.: Biological factors affecting long-term results of valvular heterografts.J. Thorac. Cardiovasc. Surg.58, 467, 1969.
5.
ClarkR.E.: Stress-strain characteristics of fresh and frozen human aortic and mitral leaflets and chordae tendinae.J. Thorac. Cardiovasc. Surg.66, 202, 1973.
6.
DavilaJ.C., MagilliganD.J.Jr., and LewisJ.W.Jr.: Is the Hancock porcine valve the best cardiac substitute today? Ann.Thorac. Surg.26, 303, 1978.
7.
De BiasiS., and PilottoF.: The ultrastructure of the leaflets of porcine aortic valves conditioned for heterografts in man.J. Submicr. Cytol.11, 353, 1979.
8.
FerransV.J., SprayT.L., BillinghamM.E., and RobertsW.C.: Structural changes in glutaraldehyde-treated porcine heterografts used as substitute cardiac valves.Am. J. Cardiol.41, 1159, 1978.
9.
HarrisP.D., KovalikA.T.W., MarksJ.A., and MalmJ.R.: Factors modifying aortic homografts structure and function.Surgery63, 45, 1968.
10.
MalmJ.R., BowmanF.O.Jr., HarrisP.D., and KovalikA.T.W.: An evaluation of aortic valve homografts sterilized by electron beam energy.J. Thorac. Cardiovasc. Surg.54, 471, 1967.
11.
PilottoF., PozzoniF., and GianiW.: A simple apparatus to evaluate the mechanical properties of small samples of animal organs.J. Med. Engineering and Technology2, 308, 1978.
12.
StinsonE.B., GrieppR.B., OyerP.E., and ShumwayN.E.: Long-term experience with porcine aortic valve xenografts.J. Thorc. Cardiovasc. Surg.73, 54, 1977.
13.
ViidikA.: Experimental evaluation of the tensile strength of isolated rabbit tendons.Bio-medical Engineering2, 64, 1967.
14.
VillaM.L., De BiasiS., and PilottoF.: Residual heteroantigenicity of glutaraldehyde-treated porcine cardiac valves.Tissue Antigens, in press.