Abstract
Background:
Thrombosis within extracorporeal membrane oxygenation (ECMO) circuits is a common complication that dominates clinical management of patients receiving mechanical circulatory support. Prior studies have identified that over 80% of circuit thrombosis can be attributed to tubing-connector junctions.
Methods:
A novel connector was designed that reduces local regions of flow stagnation at the tubing-connector junction to eliminate a primary source of ECMO circuit thrombi. To compare clotting between the novel connectors and the traditional connectors, both in vitro loops and an in vivo caprine model of long-term (48 h) ECMO were used to generate tubing-connector junction clots.
Results:
In vitro, the traditional connectors uniformly (9/9) formed large thrombi, while novel connectors formed a small thrombus in only one of nine (p < 0.0001). In the long-term goat ECMO circuits, the traditional connectors exhibited more thrombi (p < 0.04), and these thrombi were more likely to protrude into the lumen of the tubing (p < 0.001).
Conclusion:
Both in vitro and in vivo validation experiments successfully recreated circuit thrombosis and demonstrate that the adoption of novel connectors can reduce the burden of circuit thrombosis.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.