Abstract
Objective:
We aimed to elucidate the effects of the micro-structure of the pyrolytic carbon for artificial heart valves on its hydrodynamic performance.
Methods:
Bileaflet mechanical valves of GKS 23 and 29 A were randomly selected. According to ISO5840, mean transvalvular pressure (MPG), regurgitation fraction (RF), and effective orifice area (EOA) of valve were assessed. Then, parallel-groove pattern was constructed by laser etching on leaflet surface, and the valves were subjected again to the same test.
Results:
Compared with before patterning at 2, 3.5, 5, and 7 L/min, the MPG of the valves in two specifications were higher, the EOA was larger in 23 A, but smaller in 29 A, and the RF was contrary to EOA. At 5 L/min, the RF in both specifications was lower after etching at 45 bpm. At 70 bpm however, the RF in 23 A decreased, in 29 A increased.
Conclusion:
The parallel-groove pattern on leaflet surface affected the hemodynamic performance of the valve prostheses.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.