Abstract
Introduction:
Non-invasive measurement of blood pressure in patients with a left ventricular assist device (LVAD) is challenging due to the mechanical properties of these circulatory support devices. Keeping blood pressure in a target range is crucial to reduce the risk of LVAD complications. As current assessment methods require trained personnel resulting in infrequent monitoring, devices for home-based measurements are sorely needed.
Objectives:
In this paper, we report a measurement method and a custom-made monitoring device based on photoplethysmography (PPG) enabling automated measurement of the mean arterial pressure (MAP) of patients with LVAD.
Methods:
The method and the device were tested on 21 adult patients with LVAD, and the estimated MAP values were compared to MAP values measured simultaneously by a human expert applying the commonly used Doppler-based method.
Results:
Results of the proposed method showed highly significant correlation with Doppler-based MAP values (R = 0.85 for inflation, p < 0.001; R = 0.96 for deflation, p<0.001). The mean difference of the proposed method to the Doppler-based MAP values was 1.48 ± 5.43 mmHg for inflation and −0.19 ± 2.71 mmHg for deflation.
Conclusion:
The results demonstrate that the proposed method is a promising direction to achieve accurate, automated, non-invasive BP measurement, applicable in home health monitoring for patients with LVAD.
Keywords
Get full access to this article
View all access options for this article.
