Abstract
In hypothermic cardiopulmonary bypass (CPB), various vasodilators are used to control the perfusion pressure. These agents, however, often decrease the pressure excessively, and the low perfusion pressure may persist until the end of CPB. In this study we evaluate the safety and characteristics of the regulation of perfusion pressure during CPB using a volatile anesthetic, sevoflurane which has an extremely low partition coefficient. Twenty adult patients who underwent cardiac surgery were studied. Sevoflurane was applied by a vaporizer inserted into the oxygenator gas supply line. Pump flows were fixed at 2.4 L/min/m2 during the hypothermic period. Sevoflurane concentration was adjusted to keep mean arterial pressure (MAP) between 40 and 70 mmHg during CPB. Hemodynamic and metabolic parameters were measured and compared to the group we previously treated with chlorpromazine. In all cases, MAP could be maintained adequately. In the sevoflurane group, systemic vascular resistance indices (SVRI) during the rewarming period and at the end of CPB were higher, and doses of norepinephrine needed at the end of CPB were significantly lower than in the chlorpromazine group. The regulation of perfusion pressure during CPB using sevoflurane was safe and could easily maintain adequate SVRI.