Abstract
Background:
All-inside arthroscopic meniscal repairs are favored by most clinicians because of their lower complication rate and decreased morbidity compared with inside-out techniques. Until now, only 1000 cycles have been used for biomechanical testing.
Hypothesis:
All-inside meniscal repairs will show inferior biomechanical response to cyclic loading (up to 100,000 cycles) and load-to-failure testing compared with inside-out suture controls.
Study Design:
Controlled laboratory study.
Methods:
Bucket-handle tears in 72 porcine menisci were repaired using the Omnispan and Fast-Fix 360 (all-inside devices) and Orthocord 2-0 and Ultrabraid 2-0 sutures (matched controls). Initial displacement, displacement after cyclic loading (100, 500, 1000, 2000, 5000, 10,000, and 100,000 cycles) between 5 and 20 N, ultimate load to failure, and mode of failure were recorded, as well as stiffness.
Results:
Initial displacement and displacement after cyclic loading were not different between the groups. The Omnispan repair demonstrated the highest load-to-failure force (mean ± SD, 151.3 ± 21.5 N) and was significantly stronger than all the other constructs (Orthocord 2-0, 105.5 ± 20.4 N; Ultrabraid 2-0, 93.4 ± 22.5 N; Fast-Fix 360, 76.6 ± 14.2 N) (P < .0001 for all). The Orthocord vertical inside-out mattress repair was significantly stronger than the Fast-Fix 360 repair (P = .003). The Omnispan (30.8 ± 3.5 N/mm) showed significantly higher stiffness compared with the Ultrabraid 2-0 (22.9 ± 6.9 N/mm, P < .0001) and Fast-Fix 360 (23.7 ± 3.9 N/mm, P = .001). The predominant mode of failure was suture failure.
Conclusion:
All-inside meniscal devices show comparable biomechanical properties compared with inside-out suture repair in cyclic loading, even after 100,000 cycles.
Clinical Relevance:
Eight to 10 weeks of rehabilitation might not pose a problem for all repairs in this worst-case scenario.