Abstract
This laboratory-based bench study was undertaken to evaluate the accuracy and equilibration characteristics of air and saline respectively as CO2 equilibrating media in the silicone balloon of a gastric tonometer and to compare two methods of measuring air Pco2.
Two gastric tonometers were suspended in a bath containing 0.9% saline maintained at 37°C. Certified calibration gases at three different CO2 concentrations were bubbled into the bath. When the bath Pco2 measurement was stable the tonometers were primed with 5 ml of air and 2.5 ml normal saline respectively and allowed to equilibrate for 30 and 90 minutes. Following equilibration, samples were aspirated and analysed in duplicate in a blood gas analyser. Bias and precision were calculated from the measured and expected Pco2 values.
A consistent negative bias (21–23%) was seen with air at all three CO2 concentrations at 30 and 90 minutes with a coefficient of variation between 2.7 and 3.3%. Imprecise data were obtained with saline at different levels of CO2. A similar experimental set-up was used to compare air Pco2 measurement by a blood gas analyser and an infra-red analyser (Tonocap). Similar bias was obtained with the blood gas analyser with respect to air Pco2 measurement as in experiment 1. The infra-red analyser measurement was highly precise with negligible bias.
Air appears to be a better CO2 equilibration medium during bench testing of tonometry producing a systematic negative offset and requiring a uniform correction factor of 1.25. This correction factor is independent of equilibration time and equilibrating CO2 concentration. The use of the infra-red analyser eliminates any bias in the measurement of air Pco2 and obviates the need for a correction factor.
