Two disposable condenser humidifiers were evaluated in nine ventilated intensive care patients. The Portex “Humid Vent” delivered end-inspired absolute humidities of 22-26,3 g/m3 at end-inspired temperatures of 27-28.3 °C. Corresponding humidities and temperatures with the Servo “Humidifier 150” were higher and were constant, at 27.7-29 g/m3 and 29.3-29.7°C respectively. These disposable devices can be used for humidification in intensive care, but only for patients breathing room air, or on a short term basis.
HodgmanC.D., WeastR.C., ShanklandR.S., SelbyS.M.Handbook of chemistry and physics.Chem Rubber Publ Co, Cleveland1962; 2582.
2.
ChamneyA.R.Humidification requirements and techniques. Anaesthesia1969; 24: 602–614.
3.
DeryR.The evolution of heat and moisture in the respiratory tract during anaesthesia with a non-rebreathing system. Canad Anaesth Soc J1973; 20: 296–309.
4.
BurtonJ.D.K.Effects of dry anaesthetized gases on the respiratory mucous membrane. Lancet1962; 1: 235–238.
5.
ChalonJ., LoewDay, MalebrancheJ.Effects of dry anaesthetic gases on tracheobronchial ciliated epithelium. Anesthesiology1972; 37: 338–343.
6.
ForbesA.R.Humidification and mucus flow in the intubated trachea. Br J Anaesth1973; 45: 874–878.
7.
ForbesA.R.Temperature, humidity and mucus flow in the intubated trachea. Br J Anaesth1974; 46: 29–34.
8.
ShanksC.A., SaraC.A.Airway heat and humidity during endotracheal intubation 4: connotations of delivered water vapour content. Anaesth Intens Care1974; 2: 212–220.
9.
ChalonJ., TayyabM.A., RamanathanS.Cytology of respiratory epithelium as a predictor of respiratory complications after surgery. Chest1975; 67: 32–35.
10.
MarfiaS., DonahoeP.K., HendrenW.H.Effect of dry and humidified gases on the respiratory epithelium in rabbits. J Ped Surg1975; 10: 583–592.
11.
ChalonJ., PatelC., AliM., RamanathanS., CapanL., TangC.K.Humidity and the anaesthetized patient. Anesthesiology1979; 50: 195–198.
12.
NoguchiH., TakumiT., AochiO.A study of humidification in tracheostomized dogs. Br J Anaesth1973; 45: 844–847.
13.
TsudaT., NoguchiH., TakumiY., AochiO.Optimum humidification of air administered to a tracheostomy in dogs. Br J Anaesth1977; 49: 965–977.
14.
TayyabM.A., AmbiavagarM., ChalonJ.Water nebulization in a non-rebreathing system during anaesthesia. Canad Anaesth Soc J1973; 20: 728–735.
15.
ShanksC.A.Clinical anaesthesia and the multiple-gauze condenser-humidifier. Br J Anaesth1974; 46: 773–777.
16.
DeryR., PelletierJ., JaquesA., ClavetM., HoudeJ.J., Humidity in anaesthesiology III: heat and moisture patterns in the respiratory tract during anaesthesia with the semi-closed system. Canad Anaesth Soc J1967; 14: 287–292.
17.
IngelstedtS.Studies on the conditioning of air in the respiratory tract. Acta Otolaryng Suppl1956; 131: 7–80.
18.
MaplesonW.W., MorganJ.G., HillardE.K.Assessment of condenser humidifiers with special reference to a multiple-gauze model. Br Med J1963; 1: 300–305.
19.
WalkerA.K.Y., BethuneD.W.A comparative study of condenser humidifiers. Anaesthesia1976; 31: 1086–1093.
20.
RevenasB., LindholmC.E.The foam nose — a new disposable heat and moisture exchanger. Acta Anaesth Scand1979; 23: 34–39.
21.
StewardD.J.A disposable condenser humidifier for use during anaesthesia. Canad Anaesth Soc J1976; 23: 191–195.