Catheters for paediatric endotracheal suction were studied using an artificial lung with compliances in the range encountered in paediatric practice.
With various combinations of “lung” compliance, size of endotracheal tube, endotracheal connector and catheter,” tracheal” pressure and air flow between endotracheal tube and catheter were measured.
Recommendations for combinations of endotracheal tube, connector and catheter size are made. The maximum negative pressure applied to the catheter should be limited.
References
1.
BermanI. R., and StahlW. M. (1968): “Prevention of hypoxic complications during endotracheal suctioning”, Surgery, 63, 586.
2.
BoutrosA. R. (1970): “Arterial blood oxygenation during and after endotracheal suctioning in the apnoeic patient”, Anesthesiology, 32, 114.
3.
BrandstaterB., and MuallemM. (1969): “Atelectasis following tracheal suction in infants”, Anesthesiology, 31, 468.
4.
British Standards Institution (1967): “Specification for surgical suction apparatus”, Part 1, British Standard 4199.
5.
FellT., and CheneyF. W. (1971): “Prevention of hypoxia during endotracheal suction”, Ann. Surg., 174, 24.
6.
HemplemannG., HartmannW., FabelH., LeitzK. H., and NollteW. J. (1971): “Tracheobronchial suction as a problem in intensive care: continuous PO2 measurements with a polarographic micromethod during various tracheobronchial suction methods”, Prak. Anaesth. and Wied., 6, 447.
7.
MainlandJ. F. (1970): “Suction in clinical medicine”, Proceedings of the Third Asian and Australasian Congress of Anaesthesiology, 192. Butterworth, Australia.
8.
NunnJ. F., ColemanA. J., SachithanandanT., BergmanN. A., and LawsJ. W. (1965): “Hypoxaemia and atelectasis produced by forced expiration”, Brit. J. Anaesth., 37, 3.
9.
RosenN., and HilardE. K. (1960): “The use of suction in clinical medicine”, Brit. J. Anaesth., 32, 486.
10.
Van VlietP. K. J, FiskG. C., and GuptaJ. M. (1971): “Artificial ventilation in respiratory failure in the newborn”, Med. J. Aust., 2, 648.