Abstract
Aims and background
The study focused on the evaluation of the accuracy of intracranial stereotactic radiosurgery treatments delivered with helical tomotherapy by means of the InterFix™ Radiosurgery kit.
Methods and study design
Twenty-two patients received stereotactic radiosurgery treatments with single fraction dose ranging from 13 to 20 Gy depending on diagnosis. Megavoltage computed tomography scans performed prior the treatments were analyzed in order to determine the position accuracy. For 8 selected cases, they were also performed at the end of the treatment to evaluate the intra-fraction motion.
Results
Mean setup errors and standard deviations were −1.6 ± 2.2 mm, −0.2 ± 1.2 mm, 0.4 ± 1.3 mm, 0.2 ± 0.5° for the lateral (IEC-x), longitudinal (IEC-y), vertical (IEC-z) directions and rotational variation (roll), respectively. Setup error was found to be greater than 3 mm-PTV expansion in 36% of the cases. Mean intra-fraction motion was 0.5 ± 0.7 mm, −0.3 ± 0.4 mm, 0.1 ± 0.5 mm and 0.1 ± 0.2° for the IEC-x, IEC-y, IEC-z and roll, respectively.
Conclusions
Observed intra-fraction movements of less than 1 mm suggested the use of the tested fixation device for stereotactic radiosurgery treatment on helical tomotherapy providing that the image-guidance procedure is always performed prior to treatment.
Keywords
Get full access to this article
View all access options for this article.
