Abstract
Objective
To evaluate the use of ultra-low-dose computed tomography (ULDCT) for CT-guided lung biopsy versus standard-dose CT (SDCT).
Methods
CT-guided lung biopsies from 115 patients (50 ULDCT, 65 SDCT) were analyzed retrospectively. SDCT settings were 120 kVp with automatic mAs modulation. ULDCT settings were 80 kVp with fixed exposure (20 mAs). Two radiologists evaluated image quality (i.e., needle artifacts, lesion contouring, vessel recognition, visibility of interlobar fissures). Complications and histological results were also evaluated.
Results
ULDCT was considered feasible for all lung interventions, showing the same diagnostic accuracy as SDCT. Its mean total radiation dose (dose–length product) was significantly reduced to 34 mGy-cm (SDCT 426 mGy-cm). Image quality and complication rates (
Conclusions
ULDCT for CT-guided lung biopsies appears safe and accurate, with a significantly reduced radiation dose. We therefore recommend routine clinical use of ULDCT for the benefit of patients and interventionalists.
Keywords
Introduction
Computed tomography (CT)-guided percutaneous lung biopsy is routinely used to establish diagnoses, thereby also strongly influencing the therapeutic regimen and prognosis. They comprise the method of choice for both peripherally located and small lesions (>2 cm) and when bronchoscopic access is limited.1,2 Multiple CT scans (planning, guiding, and control scans) may be necessary during a single intervention, resulting in considerable radiation exposure to both patient and interventionalist.3,4 Considerable effort has therefore gone into reducing the radiation dose. Several studies5–8 have describe improved image quality in ultra-low-dose CT (ULDCT) protocols using iterative reconstruction algorithms (iDose 5, Philips Healthcare, Cleveland, OH, USA). Only limited data are available,9–11 however, regarding substantially reduced radiation dosages for ULDCT use in interventional procedures. With lung biopsies, the image quality does not have to reach a diagnostic standard, thus allowing the possibility of using ULDCT (Figure 1).
The aim of this retrospective study was to evaluate the interventional accuracy and safety of a standardized ULDCT protocol using image reconstruction via iterative reconstruction in the clinical setting.
Material and methods
Our hospital institutional review board approved this study, waiving the requirement for written informed consent because of its retrospective study design.
A total of 115 patients who had had an indication for a chest intervention were retrospectively enrolled in this study. In August 2012, we established an ULDCT protocol for lung interventions at our institution following the implementation of iterative reconstruction. The tube voltage was reduced from 120 kVp to 80 kVp, and a fixed tube current–time product of 20 mAs was used instead of the dose modulation employed for standard-dose CT (SDCT). A total of 50 patients underwent non-enhanced ULDCT of the chest for biopsy of suspicious pulmonary lesions (study group). The control group included 65 consecutive patients whose non-enhanced SDCT chest scans were reviewed. The data were collected between January 2009 and November 2014, with the interventions performed by two radiologists with more than 3 years of experience in diagnostic and interventional CT. The percutaneous biopsy procedure was standardized in our institution as follows.
A limited (planning) CT scan of the thorax (helical acquisition mode) confirms the location of the nodule and is used to determine the safest approach. The skin is disinfected, and local anesthetics are applied. A small skin incision is made at the biopsy needle entrance point. Sequential (procedural CT) images (4 × 5 mm collimation) are used to guide the needle path to the nodule. Specimens are obtained using an 18-gauge coaxial biopsy needle (Gallini Medical devices, Mantova, Italy). A postprocedural (control) CT scan (helical acquisition mode) of the thorax is used to rule out complications. A posteroanterior plain radiographic examination of the chest during expiration is performed 4 h after the intervention. The patient is discharged from hospital the following day if there are no complications.
All examinations were performed on a Brilliance iCT 256 scanner (Philips Healthcare, Cleveland, OH, USA). The detector collimation was 2 × 128 × 0.625 mm, creating 256 overlapping slices via a dynamic z-flying focal spot for the planning and control CT scans. The pitch was 0.758 and the rotation speed 0.33 s. The scan protocol for the procedural CT was sequential (axial) acquisition mode and utilized a collimation of 4 × 5 mm. The “L” (lung) filter was chosen as a reconstruction kernel. For the SDCT protocol, the dose modulation technique DoseRight, including Z-DOM (operating in the z direction), was deployed for helical scans, whereas for the procedural CT a fixed tube current–time product of 50 mAs was used. The tube voltages for the planning, procedural, and control SDCT scans were 120 kVp. The examinations were reconstructed with the iterative reconstruction algorithm iDose 5 for SDCT and ULDCT (Philips Healthcare, Cleveland, OH, USA). The reconstruction with iDose level 5 was used for the intervention itself.
The dose–length product (DLP) was recorded from the patient dose report. Histological findings were also recorded on the patients’ charts.
Important factors that may have had an influence on radiation dose and image quality were obtained from the radiological information system HIS/RIS and from the patients’ charts. Body mass index, intervention time, size of punctured lesions, and complications were evaluated. Complications were categorized using the definitions of the Society of Interventional Radiology (SIR). 12 A minor complication was defined as an event that required no further therapy or only nominal therapy—e.g., overnight admission for observation or SIR classification A (minor bleeding) or B (pneumothorax without the need of a chest tube). Mild bleeding from the puncture site was considered a regular consequence and not counted as a complication. A major complication that required therapy and minor hospitalization (<48 h) was classified as SIR C (pneumothorax with immediately placed chest tube) or SIR D (prolonged hospitalization of >48 h).
Qualitative analysis of image quality
The qualitative analysis was performed on a Picture Archiving and Communication System (Agfa Technical Imaging Systems, Richfield Park, NJ) by two radiologists with more than 3 years of experience in diagnostic/interventional CT. The radiologists worked independently and were blinded to each other’s results.
Image quality was assessed for criteria relevant to a safe biopsy, including the visual sharpness of vessels in the proposed needle pathway: 1 = sharp, 2 = limited definition but without impaired interventional confidence, 3 = limited definition with impaired interventional confidence; conspicuity of the lesion: 1 = sharp, 2 = limited definition but without impaired interventional confidence, 3 = limited definition with impaired interventional confidence; and visibility of the interlobar fissures: 1 = good, 2 = doubtful, 3 = nonexistent. Inter-reader variability was then analyzed.
Statistical analysis
Quantitative variables were summarized as the median (interquartile range) and qualitative variables as the count (percentage). Differences between groups were assessed using the Mann–Whitney U test or the
Results
Patient distribution, procedural differences, histological findings
Patient demographics of the patients in the ULDCT and SDCT groups.
ULDCT, ultra-low-dose computed tomography; SDCT, standard-dose computed tomography; BMI, body mass index.
Procedural comparison of the ULDCT and SDCT groups.
Major and minor complications in the ULDCT and SDCT groups.
SIR, Society of Interventional Radiology.
Histological findings after biopsy in the ULDCT and SDCT groups.
NSCLC, non-small cell lung cancer.
Radiation dose
Dose–length product for the ULDCT and SDCT protocols.
ULDCT and SDCT results are expressed as the median (interquartile rate) and the adjusted mean 95% confidence interval (in italics).
Expressed in mGy × cm.
Subjective image quality analysis
Qualitative image evaluation and inter-reader variability.
Vessel sharpness in the proposed needle pathway was scored as: 1 = sharp; 2 = limited definition but without impaired interventional confidence; 3 = limited definition with impaired interventional confidence.
Conspicuity of the lesion as scored as: 1 = sharp; 2 = limited definition but without impaired interventional confidence; 3 = limited definition with impaired interventional confidence.
Visibility of the interlobar fissures was scored as: 1 = good; 2 = doubtful; 3 = nonexistent.
Discussion
The use of ULDCT for CT-guided lung biopsies, which had already been implemented in routine clinical practice, was evaluated. Previous studies demonstrated that iterative reconstruction algorithms allowed the use of ULDCT while maintaining acceptable image quality.2,13 The results of our study revealed a substantial reduction in radiation exposure—reduced to 34 mGy-cm (80 kVp, 20 mAs) from 412 mGy-cm used for the SDCT intervention. This difference is a more than 10-fold reduction in the radiation dose.
When compared with previous studies,9,10 we further reduced the tube voltage (80 kV vs. 100 kV) while achieving comparable results. It proved that a dose reduction could produce the same results regarding safety, feasibility, and quality of the biopsy. A numerical comparison of the DLP, however, was not possible because of the different study designs, CT scanners, and protocols used (e.g., we routinely perform a complete chest CT scan after puncture, whereas Smith et al. scan only the needle insertion site).
With regard to image quality, we demonstrated that iterative reconstruction post-processing resulted in adequate images, although ring artifacts appeared to be common in the ULDCT group (Figure 2a and b). These artifacts mainly occurred in soft tissue masses, such as in the shoulder area or the abdomen. There were no relevant artifacts reported in the thoracic area that could have interfered with the biopsies. For one of the image quality criteria (the identification of interlobar fissures), one of the readers was unsure of the determination in 8% of cases, which might be the consequence of the lowered image quality. The other reader was positive about a fissure determination in all cases, so experience may also have to be taken into account.
In terms of the safety and feasibility of ULDCT for lung interventions, the only complication that was recorded was pneumothorax. No relevant hemorrhages or SIR D complications were recorded. No significant difference was found between the complication rates of the ULDCT and the SDCT protocols. Overall, we recorded a complication rate within the documented mean for such interventions. 14
Examination of the histological results of the biopsies revealed a high percentage of malignancy in both groups (Table 4), indicating correct needle positioning and demonstrating the importance of dose reduction: 20% of the patients had a benign diagnosis and would require further follow-up scans. Inconclusive results were found for specimens from three probes in the ULDCT group, in contrast to specimens from one probe in the SDCT group. This difference may be due to the probe sites (i.e., the specimen obtained from an area of central necrosis). In the ULDCT group, two of three patients with an inconclusive specimen showed inconspicuous clinical and radiological (CT examination) findings between 6 and 14 months after the intervention. The other patients were lost to follow-up.
A 65-year-old man (body mass index 28 g/m2) with suspected pulmonary metastases of malignant melanoma underwent ultra-low-dose computed tomography. Total radiation dose (dose–length product, DLP, 35.1 mGy × cm) with a total effective dose of 0.5 mSv (calculated by CT-Expo: Stamm G, Nagel HD. CT-Expo – a novel program for dose evaluation in CT (a, b) Ring artifacts resulting from a reduced dosage of 80 kVp and 20 mAs.

When considering our results in regard to the clinical setting, the ULDCT scan protocol for interventional lung biopsies was safe and feasible. Thus, the cumulative dose could be significantly reduced without reducing interventional accuracy compared with that achieved with SDCT (Table 5). No increased risk of complications was reported.
Summing up, our results confirm previous study results2,9,10 and take them one step farther: We could reduce the tube current while maintaining the same high procedural quality and low complication rate.
A limitation of this study is that we did not make intra-individual comparisons of the two scan protocols because it was considered ethically unjustifiable. Moreover, patients in the SDCT group were significantly older than those in the ULDCT group. In addition, the scan length differed between groups (Table 2). As these factors may affect the comparability of the radiation dose between groups, we adjusted the two groups according to age and scan length using a log-scale multivariable linear regression.
Conclusion
The use of ULDCT for performing lung interventions was shown to be feasible in all patients examined. The complication rate for the ULDCT protocol was no higher than that for the SDCT protocol, indicating that the ULDCT is a safe protocol for lung biopsy. In addition, there was a significant reduction in the cumulative radiation dose, from 412 mGy-cm to 34 mG-cm, with no reduction in the accuracy of needle placement.