Abstract
Background
The major problems of computed tomography (CT) imaging include radiation exposure and severe artifacts caused by operative implants.
Purpose
To evaluate the usefulness of the metal artifact reduction algorithm and model-based iterative reconstruction (MBIR) in postoperative low-dose (LD) spine CT.
Material and Methods
A CT torso phantom was scanned at standard-dose (SD) and LD settings. The CT images were reconstructed by three methods: hybrid iterative reconstruction (HIR); metal artifact reduction; and MBIR. The radiation dose of the phantom imaging was evaluated by volume CT dose index (mGy), dose length product (DLP, mGy × cm), and effective dose (mSv). The image quality of the six images was visually evaluated and analyzed using Scheffe’s paired comparison method. The average preference of each method was calculated based on the comparative scores. The task transfer function (TTF) and noise power spectrum for HIR and MBIR were also measured.
Results
The respective radiation-dose-related parameters of the SD and LD conditions were: volume CT dose index = 10.2 and 1.2 mGy; DLP = 277.9 and 33.9 mGy × cm; and effective dose = 4.2 and 0.5 mSv. The average preference for diagnostic acceptability of MBIR at LD was not significantly different from the other reconstructions of SD data. MBIR successfully reduced metal artifacts in the LD condition. The 10% TTF was higher for HIR at SD and higher for MBIR at LD.
Conclusion
MBIR is useful for LD spine CT after spine surgery with metal implant.
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.