Abstract
Background
Abdominal computed tomography (CT) without arm elevation is associated with degraded image quality due to streak artifacts.
Purpose
To compare the degree of streak artifacts in abdominal CT images without arm elevation between full iterative reconstruction (IR), hybrid IR, and filtered back projection (FBP) using two commercially available scanners.
Material and Methods
First, a phantom study simulating CT examination without arm elevation was performed. Second, unenhanced axial images of 33 patients (17 and 16 patients for each vendor) who underwent CT without arm elevation were reconstructed with full IR, hybrid IR and FBP. A radiologist placed 50 parallel lines with lengths of 50 pixels vertical to the streaks and quantitatively evaluated the images for streak artifacts in the phantom study. Two radiologists evaluated the images of patients for streak artifacts (on the liver and the kidney) and diagnostic acceptability using a four-point scale.
Results
The phantom study indicated that full IR algorithms were more effective than FBP in reducing streak artifacts. In the clinical patient study, streak artifacts were significantly more reduced with full IR compared with FBP in both the liver and kidney (
Conclusion
In abdominal CT without arm elevation, full IR enabled a more efficient streak artifact reduction compared with FBP and MBIR was associated with diagnostically more acceptable images.