Abstract
Background
There has been no comparison of detectability of small lung cancer between general and medical LCD monitors or no comparison of detectability of small lung cancer between solid and part-solid nodules.
Purpose
To compare the detectabilities of T1a lung cancer on chest radiographs on three LCD monitor types: 2-megapixel (MP) for general purpose (General), 2-MP for medical purpose (Medical), and 3-MP-Medical.
Material and Methods
Radiographs from forty patients with T1aN0M0 primary lung cancer (27 solid nodules, 13 part-solid nodules) and 60 patients with no abnormalities on both chest X-ray and computed tomography (CT) were consecutively collected. Five readers assessed 100 cases for each monitor. The observations were analyzed using receiver operating characteristic (ROC) analysis. A jackknife method was used for statistical analysis. A
Results
The average AUC for all T1a lung cancer nodule detection using the 2-MP-General, 2-MP-Medical, and 3-MP-Medical LCD monitors were 0.86, 0.89, and 0.89, respectively; there were no significant differences among them. The average AUC for part-solid nodule detection using a 2-MP-General, 2-MP-Medical, and 3-MP-Medical LCD monitors were 0.77, 0.86, and 0.89, respectively. There were significant differences between the 2-MP-General and 2-MP-Medical LCD monitors (
Conclusion
Detectability of part-solid nodules using a general-purpose LCD monitor was significantly lower than those using medical-purpose LCD monitors.
Get full access to this article
View all access options for this article.