Abstract
Background
Despite the well-established requirement for radiation dose reduction there are few studies examining the potential for lower extremity CT angiography (CTA) at 70 kVp.
Purpose
To compare the image quality and radiation dose of lower extremity CTA at 70 kVp using a dual-source CT system with an integrated circuit detector to similar studies at 120 kVp.
Material and Methods
A total of 62 patients underwent lower extremity CTA. Thirty-one patients were examined at 70 kVp using a second generation dual-source CT with an integrated circuit detector (70 kVp group) and 31 patients were evaluated at 120 kVp using a first generation dual-source CT (120 kVp group). The attenuation and image noise were measured and signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Two radiologists assessed image quality. Radiation dose was compared.
Results
The mean attenuation of the 70 kVp group was higher than the 120 kVp group (575 ± 149 Hounsfield units [HU]
Conclusion
Lower extremity CTA at 70 kVp allows for lower radiation dose with higher SNR, CNR, and image quality when compared with standard 120 kVp.