Abstract
Objectives
This study analyzed the maintenance of venous caliber compared to nominal package values following the deployment of venous-specific stents.
Methods
This is a retrospective review of patients undergoing iliac vein stenting with Venovo (BD, Franklin Lakes, NJ) at a single institution. Stent diameters were measured with fluoroscopy, IVUS, and 1-month post-operative duplex sonography. These measurements were compared to nominal package stent diameter.
Results
The study consists of 35 subjects with 75 stents deployed in the common iliac, external iliac, and/or the common femoral veins. Indications for treatment included post-thrombotic syndrome (57%), non-thrombotic venous obstruction (24%), and acute deep vein thrombosis (16%). For all stents, immediate post-deployment diameter, compared to nominal package diameter, changed by 5.2% by fluoroscopy (p = .01) and 4.4% by IVUS (p < .001). After deployment and venoplasty, post-deployment diameter for 16 mm stents was 14.8 mm by fluoroscopy (7.2% change; p = .004) and 15.0 mm by IVUS (6.3% change; p < .001) and 13.4 mm by fluoroscopy (4.1% change, p = .008) and 13.7 mm by IVUS for 14 mm stents (2.2% change, p = .307). Diameter at 1 month compared to immediate post-procedure stent diameter demonstrates a 5.1 % change (p = .16) and 4.4% change (p = .06).
Conclusion
Venous-specific stents demonstrate a 4.4% to 5.2% diameter decrease immediately post-deployment for 16 mm stents with no significant further reduction at 1 month. Optimal stent sizing to match expected venous diameters based on patient size is important to consider during venous reconstruction to prevent diameter reduction or stent recoil.
Get full access to this article