Abstract
The practice of flushing the cardiopulmonary bypass circuit and oxygenator with carbon dioxide (CO 2) has been shown to reduce priming time. However, the optimal flow rate of CO2 and the time for which it must be flushed through the circuit in order to remove the air, have not been determined. In this experiment, CO2 was passed through the arterial filter and arteriovenous (AV) loop at various flow rates and serial gas samples were taken from the circuit. The residual concentration of air in each sample was measured using a gas chromatograph. It was shown that as the flow rate of CO2 was increased, the time taken to remove the residual air decreased. Flushing the circuit with CO2 at a flow rate of 6 l/min for four minutes ensures a residual air concentration of less than 0.01%.