Abstract
Object:
We compared the clinical effectiveness and biocompatibility of poly-2-methoxyethyl acrylate (PMEA)-coated and heparin-coated cardiopulmonary bypass (CPB) circuits in a prospective pediatric trial.
Methods:
Infants randomly received heparin-coated (n=7) or PMEA-coated (n=7) circuits in elective pediatric cardiac surgery with CPB for ventricular septum defects. Clinical and hematologic variables, respiratory indices and hemodynamic changes were analyzed perioperatively.
Results:
Demographic and clinical variables were similar in both groups. Leukocyte counts were significantly lower 5 minutes after CPB in the PMEA group than the heparin group. Hemodynamic data showed that PMEA caused hypotension within 5 minutes of CPB. The respiratory index was significantly higher immediately after CPB and 1 hour after transfer to the intensive care unit (ICU) in the PMEA group, as were levels of C-reactive protein 24 hours after transfer to the ICU.
Conclusion:
Our study shows that PMEA-coated circuits, unlike heparin-coated circuits, cause transient leukopenia during pediatric CPB and, perhaps, systemic inflammatory respiratory syndrome after pediatric CPB.
Get full access to this article
View all access options for this article.