Abstract
We developed an attachable magnetic nerve stimulator (AMNS) that connects the metallic instruments to a neurophysiological monitoring unit for monitoring the facial nerve (FN) during ear surgery and present our experiences with intraoperative neuromonitoring (IONM) of the FN using AMNS. The FN in 20 patients who underwent tympanomastoidectomy for chronic ear disease was examined. The electromyography (EMG) amplitudes of the FN using AMNS were assessed. The EMG amplitudes of the FN at 1.00-mA stimulation during drilling were 265 ± 64 µV in the orbicularis oris and 288 ± 57 µV in the orbicularis oculi. The EMG amplitudes using AMNS attached to the surgical instruments under 0.35-mA stimulus at the tympanic segment of the FN were 196 ± 43 µV in the orbicularis oris and 197 ± 41 µV in the orbicularis oculi. The application of continuous stimulation with burr and surgical instruments using the AMNS is feasible and effective for FN stimulation and identification.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.