Abstract
Echodynamography (EDG) is a computational method to estimate and visualize two-dimensional flow velocity vectors by applying dynamic flow theories to color Doppler echocardiography. The EDG method must be validated if applied to human cardiac flow function. However, a few studies of flow estimated have compared by EDG to the flow data were acquired by other methods. In this study, EDG was validated by comparing the analysis of estimating and visualizing flow velocity vectors obtained by original particle image velocimetry (PIV) based on a left ventricular (LV) phantom hydrogel (in vitro studies) and by EDG based on the virtual Doppler velocity. Velocity measured by PIV method and velocity estimated by EDG method in the perpendicular direction and the radial direction were compared. Regression analysis for the velocity estimated in the radial direction revealed an excellent correlation (
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.