Abstract
Background: Central vein catheter position is a vital element in promoting longevity and minimizing adverse events associated with long-term parenteral nutrition. Traditionally, position has been verified using a chest radiograph. However, this mode of assessment has limitations as the catheter is placed in a dynamic system subject to forces from changes in posture and diaphragmatic movement. Methods: We compared the reported position using a chest x-ray compared with assessment using transesophageal echocardiology (TOE) in 9 patients receiving home parenteral nutrition. The x-ray was reported by a radiologist unaware of the study. Results: There were discordant results in 7 of the 9 cases with catheter tip placed in the right atrium or impinging in the tricuspid valve which was not evident from the chest x-ray. TOE offered greater information of catheter tip position and relationship to adjacent anatomy. Conclusions: Further work is required but this observational study suggests guidelines suggesting the use of a chest radiograph to confirm catheter position may need to be re-assessed. (Journal of Parenteral and Enteral Nutrition