A method is described for weighing polyvinyl chloride filters in cotton dust measurements using an electrobalance. Elaborate conditioning methods designed to overcome moisture effects are unnecessary. Filter handling and storage, balance stability, and type of sampling cassette (filter holder) are factors which affect weighing accuracy. Polystyrene cassettes should be used, and after sampling, filters contained in the cassettes should be conditioned in a standard atmo sphere and weighed immediately upon removal from the cassettes.
Get full access to this article
View all access options for this article.
References
1.
Aerosol Technology Committee, American Industrial Hygiene Association, Guide for Respirable Mass Sampling , Amer. Indust. Hygiene Assoc. J.31, 136 (1970).
2.
Jameson, T.A. and Spann, A.T., Low-Volume Micronanalytical Air Sampling, Dust Topics, 4, No. 4, Gelman Instrument Co. (Nov., 1967).
3.
OSHA Sampling Data Sheet No. 5, U. S. Dept. HEW, National Institute of Occupational Safety & Health, 1014 Broadway, Cincinnati, Ohio 45202.
4.
Vander Kolk, Alvin L., and Carter , Morgan D., Quartz Analysis of Respirable Dust by X-Ray Diffraction (Abstracts of 1971 American Industrial Hygiene Conference), Amer. Indust. Hygiene Assoc. J., 32, 15 (1971).
5.
Van Houten, R., Chew, F.F., and Lee, George, Reliability of Filters Commonly Used for Gravimetric Sampling (Abstracts of 1971 American Industrial Hygiene Conference) , Amer. Indust. Hygiene Assoc. J.32, 16 (1971).