Abstract
Background and aims
Bystanders should be protected against aerosols, droplets, saliva, blood and vomitus during resuscitation after cardiac arrest The SARUS (safer - airway - resuscitation) CPR airway hood™ is a clear plastic cover and integrated mask that envelopes the head and torso. Our objectives were to test leakage using saline aerosol generation tests, then assess the performance of the hood during mock cardio-pulmonary resuscitation on a manikin.
Methods
A checklist was validated by comparing the performance of 10 novices against 10 experts during mock resuscitation. Thereafter, 15 novices were tested with and without the hood, in a randomised cross-over study, one week apart.
Results
Laboratory analysis showed a > 99% reduction of saline particles detected 5 cm, 75 cm and 165 cm above volunteers wearing the hood. On manikins, experts scored better compared to novices, 8.5 (0.7) vs 7.6 (1.2), difference (95%CI) 0.9 (0.4–1.3), P = 0.0004. Novice performance was equivalent using the hood and standard equipment, 7.3 (1.4) vs 7.3 (1.1) respectively, difference (90%CI) 0.0 (−0.3 - 0.3), P = 0.90.
Conclusion
Aerosol transmission reduced in the breathing zone. Simulated resuscitation by novices was equivalent with and without the hood.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.