The flexure modulus of elasticity of standard stainless steel orthodontic wires was determined by the use of an iterative finite element technique to be 25.4 x 106 psi (175 x 103 MN/m2). This technique accounts for the configurational changes in the test specimens due to the relatively large deflections during the cantilever test. Under these conditions, the elementary strength of materials relationships does not accurately describe the flexure characteristics of the wires.
Get full access to this article
View all access options for this article.
References
1.
Goldberg, A.J.; Vanderby, R., Jr.; and Burstone, C.J.: Reduction in the Modulus of Elasticity in Orthodontic Wires, J Dent Res56:1227-1231,1977.
2.
Morrow, R.E.: Stored Energy in Orthodontic Wires, IADR Progr & Abst57: No. 174, 1978.
3.
Brantley, W.A.; Augat, W.S.; Myers, C.L.; and Winders, R.V.: Bending Deformation Studies of Orthodontic Wires, J Dent Res57:609-615,1978.
4.
Brantley, W.A. : Comments on Stiffness Measurements for Orthodontic Wires, J Dent Res55:705, 1976.
5.
Brantley, W.A. and Myers, C.L.: Measurement of Bending Deformation for Small Diameter Orthodontic Wires, J Dent Res58:1696-1700, 1979.
6.
Instruction Manual for Six-inch-pound Capacity Stiffness Tester: Tinius Olsen Testing Machine Co., Inc., Willow Grove, PA, p. 9.
7.
Solonche, D.J. ; Burstone, C.J.; and Vanderby , R., Jr.: A Device for Determining the Mechanical Behavior of Orthodontic Appliances, IEEE Trans on Eng in Med and Biol24:538-539, 1977.
8.
Popov, E.P.; Introduction to Mechanics of Material, Englewood Cliffs , New Jersey: Prentice-Hall, Inc., 1952, p. 435.
9.
American Dental Association Specifications No. 32 for Orthodontic Wires Not Containing Precious Metals, JADA95:1169-1171, 1977.
10.
Timoshenko, S. and Goodier, J.N.: Theory of Elasticity, 2nd ed., New York: McGraw-Hill Co., Inc., 1951, p. 39.
11.
Snedecor, G.W. and Cochran, W.G.: Statistical Methods, 6th ed., Ames, Iowa: Iowa State University Press, 1967, pp. 135-157.