Abstract
Background
Tracheotomy has been used to assist in weaning patients from mechanical ventilation. Some patients fail to be weaned from the ventilator despite tracheostomy. We hypothesized that removing the inner cannula from the tracheostomy tube would decrease the tube's imposed work of breathing (WOBIMP).
Methods
The hypothesis was tested using a lung model, by measuring the change in WOBIMP when the inner cannula was removed. A mechanical lung model was developed using a test lung to simulate a spontaneously breathing patient. WOBIMP was measured with a commercially available lung mechanics monitor. Shiley size 6, 8, and 10 nonfenes- trated tracheostomy tubes were tested with the inner cannula in and out. Breathing conditions were simulated using tidal volumes (VT) of 300 and 500 mL matched with breathing frequencies of 12, 24, and 32 breaths per minute, by using a ventilator to simulate spontaneous breathing through one side of the test lung.
Results
Under all the tested breathing conditions, WOBIMP for each of the 3 tracheostomy tubes was significantly reduced (p < 0.05) when the inner cannula was removed. Also, as simulated spontaneous inspiratory flow demand increased (ie, as VT and/or frequency were increased), WOBIMP also increased, and vice versa. With the cannula removed, WOBIMP was not significantly different between the size 6 and 8 tubes nor between the size 8 and 10 tubes when VT was 300 mL and frequency was 12 breaths per minute.
Conclusions
There was a significant decrease in WOBIMP with each tube when the inner cannula was removed. WOBIMP increased with an increase in inspiratory flow demand (ie, increase in VT and/or frequency), as well as when tube size decreased. In weaning a tracheostomized patient from mechanical ventilation, increasing the internal diameter of the tube by removing the inner cannula may be beneficial. Further study is needed to determine if these findings are clinically important.
Keywords
Get full access to this article
View all access options for this article.
