Abstract
A previously described method for mannitol in urine has been modified and improved. End product inhibition by fructose in the mannitol dehydrogenase method for mannitol has been minimized; the assay is linear over a sample mannitol concentration range of 0–12 mmol/L; no significant interference from other sugars or sugar alcohols could be demonstrated. The method is precise (within-batch CV < 1%), rapid and shows excellent recovery of mannitol in spiked samples. Comparison with gas liquid chromatography shows excellent correlation (r = 0 · 994) between the two methods.