We describe the problems encountered with the adaptation of two enzymatic assays for uric acid to a centrifugal analyser. A method employing the uricase/catalase/aldehyde dehydrogenase enzyme system sometimes suffered from interference due to light-scattering species in samples and reagents. The second method employing the uricase/peroxidase system gave results that were lower than the comparison method.
References
1.
MorgensternSFlorRVKaufmanJH, The automated determination of serum uric acid. Clin Chem.1966; 12: 748–66.
2.
FeichtmeirTVWrennHT. Direct determination of uric acid using uricase. Amer J Clin Path1955; 25: 833–9.
3.
KageyamaN. A direct colorimetric determination of uric acid in serum and urine with uricase-catalase system. Clin Chim Acta1971; 31: 421–6.
4.
HaeckelR. The use of aldehyde dehydrogenase to determine H2O2-producing reactions. (I) The determination of the uric acid concentration. J Clin Chem Clin BioChem.1976; 14: 101–7.
5.
FossatiPPrincipeLBertiG. Use of 3.5-dichloro-2-hydroxybenzene sulphonic acid/4-aminophenazone chromogenic system in direct enzymic assay of uric acid in serum and urine. Clin Chem.1980; 26: 227–31.
6.
GochmanNSchmitzJM. Automated determination of uric acid, with use of a uricase-peroxidase system. Clin Chem.1971; 17: 1154–9.
7.
WitteDLBrownLFFeldRD. Effects of bilirubin on detection of hydrogen peroxide by use of peroxidase. Clin Chem.1978; 24: 1778–82.
8.
JamesDRPriceCP. Interference in colorimetric reactions for measuring hydrogen peroxide. Ann Clin BioChem.1984; 21: 000–000.