Abstract
The fluorometric method of Fingerhut et al. (1969) for the determination of serum calcium has been modified to approximately treble the sensitivity. Using only 120 μl of sample a method has been developed for the simultaneous determination of calcium and inorganic phosphate in serum and urine, the phosphate being measured by complexing phosphomolybdate with methyl green, a method developed by Van Belle (1970). Only a single dialysis stage is required, and reagents and sample can be delivered using an AutoAnalyzer Pump I. The techniques correlate well with other methods. Accuracy and precision data are presented.