Abstract
We demonstrate the performance of a dual frequency comb quantum cascade laser (QCL) spectrometer for the application of vibrational Stark spectroscopy. Measurements performed on fluorobenzene with the dual-comb spectrometer (DCS) were compared to results obtained using a conventional Fourier transform infrared (FT-IR) instrument in terms of spectral response, parameter estimation, and signal-to-noise ratio (S/N). The dual-comb spectrometer provided similar qualitative and quantitative data as the FT-IR setup in 250 times shorter acquisition time. For fluorobenzene, the DCS measurement resulted in a more precise estimation of the fluorobenzene Stark tuning rate ((0.81 ± 0.09) cm−1/(MV/cm)) than with the FT-IR system ((0.89 ± 0.15) cm−1/(MV/cm)). Both values are in accordance with the previously reported value of 0.84 cm−1/(MV/cm). We also point to an improvement of signal-to-noise ratio in the DCS configuration. Additional characteristics of the dual-comb spectrometer applicable to vibrational Stark spectroscopy and their scaling properties for future applications are discussed.
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.