Abstract
Objectives:
To determine the characteristics of laser-protected endotracheal tubes (LPETs) valued by otolaryngologists performing transoral laser surgery in the head and neck and to measure LPET stiffness.
Methods:
An online questionnaire was completed by American Broncho-Esophagological Association (ABEA) and American Head and Neck Society (AHNS) members. LPET distal end compliance was measured in a biomechanics laboratory.
Results:
A total of 228 out of 2109 combined ABEA and AHNS members completed the survey. The following LPET characteristics, which were properties of the Medtronic Laser-Shield II tube (MLST), were highly valued: softness and flexibility, surface smoothness, and a tight-to-shaft balloon (all P < .01). Prior to industry-driven discontinuation of the MLST, 52% of surgeons (78% of fellowship-trained laryngologists [FTLs]) reported using it; afterward, 58% reported using the stainless steel, Mallinckrodt Laser-Flex tube (MLFT). Forty-six percent of all respondents (69% of FTLs) did not consider cost being a factor in LPET choice. Biomechanical testing revealed the distal end of the MLST to be 3.45 times more compliant than the MLFT (P < .01).
Conclusion:
Members of the ABEA and AHNS, particularly FTLs, highly value distinguishing properties of the now discontinued MLST. Manufacturers should consider this in the design of new LPETs.
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.