Abstract
To determine the optimal air stimulus intensity and duration for elicitation of the laryngo–upper esophageal sphincter (UES) contractile reflex, we studied 37 healthy volunteers 20 to 81 years of age. A sleeve device monitored the UES pressure. For laryngeal stimulation, we used an air stimulator unit (Pentax AP-4000) that incorporated a nasolaryngeal endoscope. The arytenoids and interarytenoid areas were stimulated at least three times by three different stimuli: 6–mm Hg air pulse with 50-ms duration, 10–mm Hg air pulse with 50-ms duration, and 6–mm Hg air pulse with 2-second duration. Of 1,165 air stimulations, 1,041 resulted in mucosal deflections. Of these, 451 resulted in an abrupt increase in UES pressure. The response/deflection ratio for 6–mm Hg stimulation with 2-second duration was significantly higher than those for air pulses with 50-ms duration (p < .001). We conclude that although the laryngo-UES contractile reflex can be elicited by an air pulse with 50-ms duration, this ultrashort stimulation is not reliable. Using longer-duration pulses (at least 2 seconds) improves the reliability of elicitation of the laryngo-UES contractile reflex.