Abstract
Objective
There are many commercially available instruments for measuring electrical conductance, but there is little information about their reliability. The aim of this study was to quantify measurement variability and assess reliability of the AcuGraph system—a commonly used electrodermal screening device.
Methods
Four experiments were conducted to measure variability in electrical conductance readings obtained by the AcuGraph system. The first involved measuring known resistors. The second measured non-human organic matter. The third was a test–retest assessment of the
Results
Against known resistors, the AcuGraph had an average coefficient of variability (CV) of 1.8% between operators and test–retests. On non-human organic material the AcuGraph had an average CV of 0.9% and 2.8%. When a single operator tested 30 participants, the average reliability for the
Conclusions
The AcuGraph measures known resistors and organic matter accurately and reliably. Skin conductance at acupoints recorded by one operator was also reliable. There was less consistency in electrodermal recordings obtained by seven different operators. Operator training and technical improvements to the AcuGraph may improve consistency among operators.