Abstract
Background:
We investigated the ability of a novel stand-alone, smartphone-based system, the cvrPhone, in estimating the minute ventilation (MV) from body surface electrocardiographic (ECG) signals.
Methods:
Twelve lead ECG signals were collected from anesthetized and mechanically ventilated swine (n = 9) using standard surface electrodes and the cvrPhone. The tidal volume delivered to the animals was varied between 0, 250, 500, and 750 mL at respiration rates of 6 and 14 breaths/min. MV estimates were determined by the cvrPhone and were compared with the delivered ones.
Results:
The median relative estimation errors were 17%, −4%, 35%, −3%, −9%, and 1%, for true MVs of 1,500, 3,000, 3,500, 4,500, 7,000, and 10,500 breaths*mL/min, respectively. The MV estimates at each of the settings were significantly different from each other (p < 0.05).
Conclusions:
We have demonstrated that accurate MV estimations can be derived from standard body surface ECG signals, using a smartphone.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.