Abstract
For chronic myelocytic leukemia patients with very high numbers of platelets, we describe an efficient method for the collection of peripheral blood stem cells (PBSC) using the Fresenius AS104 cell separator. In these patients, it is difficult to collect a sufficient number of PBSC, due to the platelet band interfering with the machine's red cell interface sensor. We, therefore, tried a manual adjustment of the device. The collection phase was set automatically. When the whole blood began to separate into the red cell layer and plasma (plus mononuclear cell) layer, the red cell interface setting of "7:1" was changed to "OFF," and the plasma pump flow rate was controlled manually in order to locate the interface position 1 cm from the outside wall of the centrifuge chamber. After the collection phase, the procedure was returned to the automatic setting. By repeating this procedure, we were able to collect large numbers of PBSC.