Abstract
Objective:
The aim of this study was to perform a systematic and multifaceted comparison of thermal effects during soft tissue ablation with various lasers and an electroscalpel (ES).
Materials and methods:
Er:YAG, Er,Cr:YSGG, CO2, Diode, Nd:YAG lasers (1 W, pulsed or continuous wave), an ES, and a scalpel (Sc; control), were employed for porcine gingival tissue ablation. Temperature changes during ablation were measured by using an infrared thermal imaging camera and a thermocouple. After ablations, the wounds were observed using stereomicroscopy and scanning electron microscopy (SEM), and histological sections were analyzed. Compositional analysis was also performed on ablated sites by SEM wavelength dispersive X-ray spectroscopy.
Results:
The surface temperature during irradiation was highest with CO2 (over 500°C), followed by Diode (267°C) and Nd:YAG (258°C), Er:YAG (164°C), ES (135°C), and Er,Cr:YSGG (85°C). Carbonization was negligible (Er:YAG), slight (Er,Cr:YSGG), moderate (Nd:YAG and ES), and severe (CO2 and Diode). Under SEM observation, Er:YAG and Er,Cr:YSGG showed smooth surfaces but other devices resulted in rough appearances. Histologically, the coagulated and thermally affected layer was extremely minimal (38 μm in thickness) and free from epithelial collapse for Er:YAG. Compared with other devices, less compositional surface change was detected with Er:YAG and Er,Cr:YSGG; additionally, the use of water spray further minimized thermal influence.
Conclusions:
Among various power devices, Er:YAG laser showed the most efficient and refined gingival ablation with minimal thermal influence on the surrounding tissues. Er:YAG and Er,Cr:YSGG lasers with water spray could be considered as minimally invasive power devices for soft tissue surgery.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.