Abstract
Colistin HiMIC Plate Kit (HiMedia Laboratories), a new commercial broth microdilution (BMD) test for colistin susceptibility testing was evaluated. BMD according to ISO standard 20776–1 (2019) with two-fold dilutions from 128 to 0.125 mg/L was used as a reference method. The colistin reference MICs (minimal inhibitory concentration) ranged from 0,25 to 128 mg/L with 15 (20.5%; 15/73) isolates having colistin reference MICs close to the current EUCAST breakpoint (MICs of 2, 4, and 8 mg/L). The study assessed the compliance of a commercial kit with the CLSI criteria, including categorical agreement (CA) and essential agreement (EA ≥90%), very major error (VME rate) <3%, and major error (ME) rate <3%. On 73 carbapenemase-producing Klebsiella pneumoniae isolates Colistin HiMICTM Plate Kit showed CA and EA of 100% (73/73; 95% CI: 0.97–1.00) and 82.2% (60/73; 95% CI: 0.72–0.90), respectively. No ME (false-resistant results) and VME (false-susceptible results) were detected. Kit showed acceptable CA, ME, and VME error parameters, whereas the EA did not meet the ≥90% threshold. Laboratories must check for possible limitations of commercial kits before they can be used for colistin susceptibility testing.