Abstract
Abstract
Background:
Pediatric robot-assisted surgery is increasingly being performed, but it is difficult to perform this procedure in infants. A pediatric thoracoscopic model of a 1-year-old patient was developed in our previous study, and this model was used to evaluate the use of a surgical robot for infant surgery.
Methods:
Eight pediatric surgeons performed an intracorporeal suturing and knot-tying task using the da Vinci Xi Robotic Surgical System. The task completion time, number of needle manipulations, and force applied during suturing of the robot-assisted thoracoscopic surgery (RATS) group were compared with those of the video-assisted thoracoscopic surgery (VATS) group whose data had been collected from the same 8 surgeons in our previous study.
Results:
The RATS group showed a significantly shorter completion time than the VATS group in the knot-tying phase (P = .016) and in the total phase (P = .0078). The RATS group showed a significantly smaller number of manipulations than the VATS group in the total phase (P = .039). The RATS group showed a significantly smaller pushing force index than the VATS group in the suturing phase (P = .031), knot-tying phase (P = .031), and in the total phase (P = .031). A seventh rib in the model was dislocated in all RATS group cases.
Conclusions:
The da Vinci Surgical System might be useful in infants because of fast movement and small pushing force. However, the robotic 8 mm instruments were too large for use in the thoracic cavity of the 1-year-old infant.