Abstract
Abstract
Background:
A validated method to predict lung deposition for inhaled medication from in vitro data is lacking in spite of many attempts to correlate in vitro and in vivo outcomes. By using an in vivo-like in vitro setup and analyzing inhalers from the same batches, both in vitro and in vivo, we wanted to create a situation where information from the in vitro and in vivo outcomes could be analyzed at the same time.
Method:
Nine inhalation products containing either budesonide or AZD4818 were evaluated. These comprised two pressurized metered dose inhalers (pMDIs), a pMDI plus a spacer, four dry powder inhalers, and two dosimetric nebulizers. In vitro, an in vivo-like setup consisting of anatomically correct inlet throats were linked to a flow system that could replay actual inhalation flow profiles through the throat to a filter or to an impactor. In vivo, total lung deposition was measured in healthy adults by pharmacokinetic methods.
Results and Conclusion